Tonometry is the measurement of tension or pressure [1]. A tonometer is an instrument for measuring tension or pressure [2].

In ophthalmology, tonometry is the procedure eye care professionals perform to determine the intraocular pressure (IOP), the fluid pressure inside the eye. Ophthalmology is the branch of Medicine which deals with the diseases and surgery of the visual pathways including the Eye, Brain An eye care professional is an individual who provides a service related to the Eyes or vision. Intraocular pressure (IOP is the fluid Pressure inside the Eye. Pressure (symbol 'p' is the force per unit Area applied to an object in a direction perpendicular to the surface Eyes are organs that detect Light, and send signals along the Optic nerve to the visual areas of the brain It is an important test in the evaluation of ocular conditions such as glaucoma as well as conditions such as phthisis bulbi, and iritis. Glaucoma is a group of diseases of the Optic nerve involving loss of retinal ganglion cells in a characteristic pattern of optic neuropathy. Phthisis bulbi is a shrunken non-functional Eye that results from severe Ocular disease, Inflammation, or injury [http//insight Iritis is a form of anterior Uveitis and refers to the Inflammation of the iris of the Eye. Most tonometers are calibrated to measure pressure in mmHg. The torr (symbol Torr) is a non- SI unit of Pressure defined as 1/760 of an atmosphere.

Contents

Methods of ophthalmotonometry

The slit lamp of a Goldmann tonometer
The slit lamp of a Goldmann tonometer
Semicircles seen during Goldmann tonometry through slit lamp
Semicircles seen during Goldmann tonometry through slit lamp
The PASCAL® Dynamic Contour Tonometer
The PASCAL® Dynamic Contour Tonometer

Position of the Transpalpebral Diaton Tonometer Tip on the Eyelid with No Contact with Cornea or use of Anesthesia
Position of the Transpalpebral Diaton Tonometer Tip on the Eyelid with No Contact with Cornea or use of Anesthesia
Hand-held Diaton tonometer with Test and Training Eye case
Hand-held Diaton tonometer with Test and Training Eye case

In Applanation tonometry, a special calibrated disinfected probe attached to a slit lamp biomicroscope is used to flatten the central cornea a fixed amount. The slit lamp is an instrument consisting of a high-intensity Light source that can be focused to shine as a slit The cornea is the transparent front part of the Eye that covers the iris, Pupil, and Anterior chamber. Because the probe makes contact with the cornea, a topical anesthetic, such as oxybuprocaine, tetracaine, alcaine, proxymetacaine or proparacaine is introduced onto the surface of the eye in the form of one or a few eye drops. A topical anesthetic is a Local anesthetic that is used to numb the surface of a body part Oxybuprocaine ( INN) also known as benoxinate, is the name of a Local anesthetic, which is used especially in Ophthalmology and Otolaryngology Tetracaine ( INN, also known as amethocaine; trade name Pontocaine) is a potent Local anesthetic of the Ester group Alcaine is a municipality located in the province of Teruel, Aragon, Spain. A topical anesthetic is a Local anesthetic that is used to numb the surface of a body part Proparacaine (Alcaine Ak-Taine is a Topical anesthetic drug of the amino ester group Eye drops are saline -containing Drops used as a vector to administer medication in the Eye. A yellow fluorescein dye is used in conjunction with a cobalt blue filter to aid the examiner in determining the IOP. Fluorescein is a Fluorophore commonly used in Microscopy, in a type of Dye laser as the Gain medium, in Forensics and Serology

  • Goldmann tonometry is considered to be the gold standard in tonometry as it is the most widely accepted method of determining approximate intraocular pressure. In Medicine, a gold standard test or criterion standard test is a Diagnostic test or benchmark that is regarded as definitive [2][3] However, as James D. Brandt, MD, points out, Goldmann tonometry is an inherently imprecise measurement.

Dynamic contour tonometry (DCT) is a novel measuring technique using the principle of contour matching instead of applanation to eliminate the systematic errors inherent in previous tonometers. These factors include the influence of corneal thickness, rigidity, curvature, and elastic properties. It is not influenced by mechanical changes, such as those seen in refractive surgery that would otherwise cause error in applanation tonometers. The PASCAL Dynamic Contour Tonometer (Ziemer Ophthalmics) is a relatively new device that uses DCT to measure IOP. Blaise Pascal (blɛz paskal (June 19 1623 &ndash August 19 1662 was a French Mathematician, Physicist, and religious Philosopher Although this device is similar in appearance to a Goldmann, the PASCAL it is unlike Goldmann applanation in that it is not a variable force tonometer.

PASCAL uses a miniature pressure sensor embedded within a tonometer tip contour-matched to the shape of the cornea. A pressure sensor measures the Pressure, typically of Gases or Liquids. The tonometer tip rests on the cornea with a constant appositional force of one gram. This is an important difference from all forms of applanation tonometry in which the probe force is variable. When the sensor is subjected to a change in pressure, the electrical resistance is altered and the PASCAL's computer calculates a change in pressure in concordance with the change in resistance.

The contour matched tip has a concave surface of radius 10. 5 mm, which approximates the cornea’s shape when the pressures on both sides of it are equal. This is the key to the PASCAL’s ability to neutralize the effect of intra-individual variation in corneal properties. Once a portion of the central cornea has taken up the shape of the tip, the integrated piezoresistive pressure sensor begins to acquire data, measuring IOP 100 times per second. The piezoresistive effect describes the changing Electrical resistance of a material due to applied Mechanical stress. A pressure sensor measures the Pressure, typically of Gases or Liquids. A complete measurement cycle requires about 8 seconds of contact time. During the measurement cycle, audio feedback is generated, which helps the clinician insure proper contact with the cornea. Literature references: [4][5][6]

Diaton tonometer measures intraocular pressure through the Eyelid. An eyelid is a thin fold of skin that covers and protects an Eye. It is regarded as a simple and safe method of ophthalmotonometry. Transpalpebral tonometry requires no contact with the cornea, therefore sterilization of the device and topical anesthetic drops are not required and there is very little risk of infection. The cornea is the transparent front part of the Eye that covers the iris, Pupil, and Anterior chamber. Sterilization (or sterilisation, see spelling differences) refers to any process that effectively kills or eliminates transmissible agents (such as Fungi A topical anesthetic is a Local anesthetic that is used to numb the surface of a body part Compared with the current standard of Goldmann applanation tonometry and Tono-Pen, recent data suggests that transpalpebral tonometry via the Diaton tonometer is clinically useful device for measuring IOP in routine eye exams. Eyes are organs that detect Light, and send signals along the Optic nerve to the visual areas of the brain [7] [8] [9] Although requiring practice of the technique [10], transpalpebral tonometery has been described as sufficient for clinical use and it may have applications in the home-monitoring of intraocular pressures. [11] Use of transpalpebral tonometry may be indicated in those for whom Goldmann tonometry is not indicated[12], such as in children, those with corneal pathology, or those who have had corneal surgery. [2][13]

Thus, corneal thickness is an important factor of IOP evaluation and monitoring, it requires the necessity of including corneal pachymetry in the program of examination the patients with suspicion of glaucoma and hypertension, especially after various keratorefractive surgeries while using the traditional corneal methods of ophthalmotonometry. Glaucoma is a group of diseases of the Optic nerve involving loss of retinal ganglion cells in a characteristic pattern of optic neuropathy. Hypertension, also referred to as high blood pressure, HTN or HPN, is a medical condition in which the Blood pressure is chronically elevated At the same time clinical application of trans-palpebral and trans-scleral Diaton tonometer makes it possible to evaluate IOP using only one device, the procedure being efficient, economical, simple in performance and requiring no additional instrumental examination.

The images to the right and below show painlessness of the Diaton tonometry procedure and its proper positioning on the Eyelid, where the edge of the lid coincides with the edge of the limbus. An eyelid is a thin fold of skin that covers and protects an Eye. Since the screening occurs above the cornea, contact lenses are not being taken out. The cornea is the transparent front part of the Eye that covers the iris, Pupil, and Anterior chamber. A contact lens (also known simply as a contact) is a corrective, cosmetic, or therapeutic lens usually placed on the Cornea

Rebound tonometry (such as iCare Tonometer) is a new method for measuring intra-ocular pressure. Unique and carefully studied and tested measurement technique which requires no local anaesthesia. An ultralight, sensitive probe touches the eye for a moment, and does not even always cause, for example, the corneal reflex (blinking). The disposable probe is also safe for the eye in microbiological terms. The accurate measurement device is primarily intended for professional use by general practitioners, opticians, nurses.

It is the only portable device of its kind that does not require anaesthesia. Compact and easy to use, it is suited to the investigation of patients such as children. Made of durable plastic and anodized aluminium, the device is light and small. Comfort and ease of use in different positions have been taken into account. The device employs an electromagnetic field, and the disposable probe is easy to change. It is also sturdy and the possible movement of the patient during investigation will not affect it.

  • Schiötz tonometry is a type of impression tonometery, historically was used to determine IOP. This type of tonometry makes use of a plunger to indent the cornea. The IOP is determined by correlation of scale reading using a nomogram, with additional small metal weights added for higher levels of IOP. This article is about the graphical devices called nomograms For a description of the Japanese puzzle see Nonogram.
  • Perkins tonometer is a special type of portable applanation tonometer, which allows measurement of IOP in children, patients unable to cooperate for slit lamp exam, and in anesthetised patients who may be in a supine position.
  • MacKay Marg tonometer
  • Palpation, also known as digital tonometry, to measure intraocular pressure is performed by gently pressing the fingertips of both index fingers onto the upper part of the bulbus through the eyelid. [13]
  • Ocular Response Analyzer The Ocular Response Analyzer (ORA, Reichert, Inc) utilizes non-contact tonometer (air puff) technology. However, in the ORA, the air pulse causes the cornea to go beyond applanation, into an indented state, before returning to normal curvature. As such, the cornea passes through applanation twice during the process, once on the way inward, and again on the way back out. Viscous damping (energy absorption) in the corneal tissue results in two different pressure values from the inward and outward applanation events. The difference between these two pressure values is termed "corneal hysteresis" and is a measurement of the cornea's biomechanical tissue properties. The ability of the ORA to measure these properties enables the calculation of an IOP measurement called IOPcc (Corneal Compensated IOP). This is an IOP measurement that is less influenced by corneal properties such as resistance or thickness. As such, it has been shown to be more accurate than other methods of tonometry. [14]

Gallery

References

  1. ^ a b c Cline D; Hofstetter HW; Griffin JR. Dictionary of Visual Science. 4th ed. Butterworth-Heinemann, Boston 1997. ISBN 0-7506-9895-0
  2. ^ a b Amm M, Hedderich J. "[Transpalpebral tonometry with a digital tonometer in healthy eyes and after penetrating keratoplasty. ]" Ophthalmologe. 2005 Jan;102(1):70-6. PMID 15322801.
  3. ^ Schlote T, Landenberger H. "[Intraocular pressure difference in Goldmann applanation tonometry versus a transpalpebral tonometer TGDc-01'PRA' in glaucoma patients]. " Klin Monatsbl Augenheilkd. 2005 Feb;222(2):123-31. PMID 15719316.
  4. ^ Kaufmann C, Bachmann LM, Thiel M, Comparison of Dynamic Contour Tonometry with Goldmann Applanation Tonometry. Invest Ophthalmol and Vis Sci 2004, Vol. 45, No. 9, 3118-3121. PMID 15326129
  5. ^ Kaufmann C, Bachmann LM, Thiel MA, Intraocular Pressure Measurements Using Dynamic Contour Tonometry after Laser In Situ Keratomileusis. Invest Ophthalmol and Vis Sci 2003;44:3790-3794. PMID 12939293
  6. ^ Kniestedt C, Nee M, Stamper RL, Dynamic Contour Tonometry. A Comparative Study on Human Cadaver Eyes. Arch Ophthalmol. 2004; 122:1287-1293. PMID 15364707
  7. ^ R. S. Davidson; N. Faberowski ; R. J. Noecker ; M. Y. Kahook " Comparison of the Diaton Transpalpebral Tonometer Versus Goldmann Applanation " ASCRS/ASOA2007 May; Poster # P-130
  8. ^ Theodore H. Curtis, M. D. 1, Douglas L Mackenzie, M. D. 1, Robert J. Noecker M. D. 2, and Malik Y. Kahook M. D. "Comparison of the Diaton Transpalpebral Tonometer Versus Tono-Pen Applanation" ASCRS/ASOA2007 May; Poster # P-128
  9. ^ Lam AK, Lam CH, Chan R. "The validity of a digital eyelid tonometer (TGDc-01) and its comparison with Goldmann applanation tonometry - a pilot study. " Ophthalmic Physiol Opt. 2005 May;25(3):205-10. PMID 15854065.
  10. ^ Henry D. Perry, M. D. , "Eyeworld magazine COVER FEATURE What’s ahead in 2007 Decking the halls with new products from 2006" Eyeworld Magazine2006 ;
  11. ^ Nesterov AP, Piletskii GK, Piletskii NG. "[Transpalpebral tonometer for measuring intraocular pressure. ]" Vestn Oftalmol. 2003 Jan-Feb;119(1):3-5. PMID 12608032.
  12. ^ Sandner D, Bohm A, Kostov S, Pillunat L. "Measurement of the intraocular pressure with the 'transpalpebral tonometer' TGDc-01 in comparison with applanation tonometry. " Graefes Arch Clin Exp Ophthalmol. 2005 Jun;243(6):563-9. Epub 2004 Dec 22. PMID 15614540.
  13. ^ a b Troost A, Yun SH, Specht K, Krummenauer F, Schwenn O. "Transpalpebral tonometry: reliability and comparison with Goldmann applanation tonometry and palpation in healthy volunteers." Br J Ophthalmol. 2005 Mar;89(3):280-3. PMID 15722303.
  14. ^ Felipe A. Medeiros, MD and Robert N. Weinreb, MD. "[Evaluation of the Influence of Corneal Biomechanical Properties on Intraocular Pressure Measurements Using the Ocular Response Analyzer]. " J Glaucoma Volume 15, Number 5, October 2006.